DESCRIPTION OF OG CARE - Blood glucose meters
OGCare ® is a simple and accurate system for determining the approximate concentration of glucose in the blood. It is most useful element of home blood glucose monitoring for people with diabetes mellitus or hypoglycemia.
OG CARE® blood glucose meter uses the amperometric method which means it measures blood glucose level electric current generated at a specific moment in time. OG CARE® blood glucose meter meets accuracy standards set by the International Organization for Standardization (ISO).
The key principle how it works:
A small drop of blood, achieved by pricking the skin with a lancet or something sharp, is placed on a disposable test strip and the meter reads and uses it to calculate the blood glucose level of a user. The meter then displays the level in units of mg/dl or mmol/l.
The user of these blood glucose meters should be aware that variety of factors can affect the accuracy of a test. These factors include calibration of a meter, ambient temperature, pressure used to wipe off strip, size and quality of blood sample, high levels of certain substances (such as ascorbic acid) in patient’s blood, hematocrit, dirt found on meter, humidity, and aging of test strips.
OG Care ® blood glucose meter main advantages:
• Auto-coding
• Reduced blood volume 0.5 µl
• Reading time just 5 seconds
• Easy to use
• Wide display
• Strips ejector
Technical features of OG CARE® blood glucose meter:
• Measurement Range: 20-60 mg/dl (1.1 - 33.3 mmol/L)
• 2X3V CR 2032 Battery. Can be used for approx 1000 measurements
• LCD display
• Memory: 300 tests result including time and date
• Dimensions: length 88 mm; height 51 mm; width 18 mm
• Weight: 48g
• USB Connection Cable
• Meter box consists of: 1 device, 1 lancing device, 1 bag of 10 pcs sterile lancets, 1 vial of 10 pcs glucose strips, 1 softcase, 1 user’s manual, 1 instruction of use for lancing device
• 3 years warranty
Technical features of OG CARE® glucose strips:
• Amperometric method of measurement
• Measurement Range: 20 - 600 mg/dl (1.1 - 33.3 mmol/L)
• Reading time: 5 seconds
• Sample Volume: 0.5µl
• Hematocrit Range: 20 - 60%
• Packaging: boxes of 25, 50 pcs
Follow the link to find more information about OG CARE® blood glucose meter at Biochemical Systems International: http://www.biosys.it/en/pr_ogcare.html
SPECIFICATION OF OG CARE - Blood glucose meters
Measurement method: | non-invasive |